Study of a lead glass calorimeter with vacuum phototriode read-out

https://doi.org/10.1016/0168-9002(86)91015-6Get rights and content

Abstract

A prototype calorimeter of lead glass bars read by one-stage vacuum photomultipliers (triodes) has been tested at CERN with electrons in the energy range 2–30 GeV. Different glass-triode combinations have been compared. An electronic amplification noise corresponding to about 30 MeV has been obtained. The space resolution has been measured to be 4.4 mm on average.

References (7)

  • B. Powell

    Nucl. Instr. and Meth.

    (1982)
  • DELPHI Progress Report, CERN/LEPC...
  • M.D. Rousseau

    IEEE Trans. Nucl. Sci. NS30

    (1983)
There are more references available in the full text version of this article.

Cited by (0)