Gas chromatography
;
Capillary, glass
;
Two-dimensional technique
;
Critical evaluation of individual instrumental components
;
Minimization of extra column effects and operational parameters discussed
;
Chemistry
;
Analytical Chemistry and Spectroscopy
Source:
Wiley InterScience Backfile Collection 1832-2000
Topics:
Chemistry and Pharmacology
Notes:
A versatile two-dimensional gas chromatograph is described, consisting of 2 separate ovens, one intermediate trap, an auxiliary inlet, and the necessary hardware to effect off-line switching according to the principle of Deans. The unit has been designed for use with high resolution glass capillary columns. The performance of individual instrumental components was critically evaluated. Results showed that low dead volume glass to metal connections were required in the manifold and detector lines to minimize extra-column effects. The mass of the intermediate trap must be low to allow rapid heating. Operational parameters are discussed and examples of some applications are shown.
Additional Material:
4 Ill.