ISSN:
0935-6304
Keywords:
Gas chromatrography, GC
;
Capillary columns
;
Splitless and on-column injection
;
Interlaboratory study
;
Chlorobiphenyl congeners
;
Internal standards
;
Chemistry
;
Analytical Chemistry and Spectroscopy
Source:
Wiley InterScience Backfile Collection 1832-2000
Topics:
Chemistry and Pharmacology
Notes:
The Community Bureau of Reference has organized a collaborative interlaboratory project to improve the analytical protocol for some specific chlorobiphenyls (CBs) within the European Community (EC). A series of test procedures were prescribed to optimize the gas chromatographic conditions for splitless and on-column injection, which substantially improved the quality of data. Important parameters included the initial column and injector temperatures and the choice of the polarity of the stationary phase. In a study with cleaned eel-fat extracts, coefficients of variation for reproducibility CV(R) ranging from 11 to 24% at the 0.04-0.3 mg/kg level per CB congener were obtained. No significant difference could be found between splitless- and on-column injection.
Additional Material:
4 Tab.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1002/jhrc.1240080824