Library

Your email was sent successfully. Check your inbox.

An error occurred while sending the email. Please try again.

Proceed reservation?

Export
  • 1
    Electronic Resource
    Electronic Resource
    Weinheim : Wiley-Blackwell
    Journal of High Resolution Chromatography 17 (1994), S. 695-699 
    ISSN: 0935-6304
    Keywords: High resolution gas chromatography ; Sulfur chemiluminescence detection ; Chromatographic injection profile ; Gaseous injection ; Sulfur gases ; Chemistry ; Analytical Chemistry and Spectroscopy
    Source: Wiley InterScience Backfile Collection 1832-2000
    Topics: Chemistry and Pharmacology
    Notes: A one meter stainless steel capillary column coated with a 20μm film of polydimethylsiloxane was used to replace the Teflon gas sampling loop of a HRGC - flameless SCD system. This column trap was equilibrated with the gaseous sample and then purged with carrier gas while being sequentially, resistively heated in ten 10 cm sections. The sequential heating of the 10cm sections was timed such that the velocity of the moving heated zone was approximately the same as the velocity of the carrier gas. The thick stationary phase film served to retard the front end of the injection band while the ballistic heating accelerated the back end. The net result was focusing of the injection, which gave rise to a narrower injection profile than that obtained with the original Teflon loop.
    Additional Material: 10 Ill.