ISSN:
0935-6304
Keywords:
Capillary gas chromatography
;
Film thickness
;
Pressure drop
;
Speed of analysis
;
Applications
;
Chemistry
;
Analytical Chemistry and Spectroscopy
Source:
Wiley InterScience Backfile Collection 1832-2000
Topics:
Chemistry and Pharmacology
Notes:
The chromatographic performance of 50 μm internal diameter (i.d.) fused silica columns coated with up to 2 μm films of immobilized SE-54 (methyl phenyl (5%) silicone) is evaluated under gas chromatographic conditions. The influence of pressure drop on the plate height is discussed. In comparison to thick film 250-530 μm i.d. columns, much higher efficiencies and faster analyses are obtained. Practical examples, performed on a standard GC instrument, illustrate the features of thick film 50 μm i.d. columns.
Additional Material:
10 Ill.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1002/jhrc.1240180202