    Rapid Communications in Mass Spectrometry 8 (1994), S. 781-785 
    ISSN: 0951-4198
    Keywords: Chemistry ; Analytical Chemistry and Spectroscopy
    Source: Wiley InterScience Backfile Collection 1832-2000
    Topics: Physics
    Notes: The ability to simultaneously focus a wide mass range of metastable fragment ions formed after the initial ionization event in a matrix-assisted laser desorption/ionization time-of-flight mass spectrometer has been made possible by the development of a new type of ion reflector. This coaxially designed time-of-flight instrument employs a modified single stage reflector whose axial voltage gradient rises differentially in order to produce an alignment of energy focal points for product ions. In contrast, product-ion focusing in conventional constant field reflectors occurs over a broad range of distances from the reflectron exit. Approximately 90% of the product-ion mass spectrum can be collected without adjustment, thereby eliminating the need to scan the reflector voltage. Design considerations of the curved field reflectron, its calibration properties and representative metastable spectra of several peptides are discussed.
    Additional Material: 6 Ill.