    Chichester : Wiley-Blackwell
    Biological Mass Spectrometry 14 (1987), S. 257-261 
    ISSN: 0887-6134
    Keywords: Chemistry ; Analytical Chemistry and Spectroscopy
    Source: Wiley InterScience Backfile Collection 1832-2000
    Topics: Chemistry and Pharmacology
    Notes: This study identified the reason for the poor quantification of Iormetazepam-TMS (1) using negative ion chemical ionization with lormetazepam-1,1,1-2H3-TMS (2) as an internal standard. Mass spectra of lormetazepam and its deuterium labelled compounds determined at various ion source temperatures (100-250°C) gave almost the same behaviour for 1 and lormetazepam-3′,4′,5′,6′-2H4-TMS (3) but a different one for 2, suggesting that the poor quantification was due to an isotope effect. This was confirmed by the findings that the ratios of ion currents of the base peaks of 1 and 3 were independent of the ion source temperature but those of 1 and 2 varied markedly with it. This phenomenon was also observed in the mass fragmentography of the molecular ion, although to a lesser degree than that of the above fragment. In both positive ion chemical ionization and electron impact ionization modes, no isotope effect arose because there was no corresponding fragment to cause the isotope effect.
    Additional Material: 9 Ill.