Library

Your email was sent successfully. Check your inbox.

An error occurred while sending the email. Please try again.

Proceed reservation?

Export
  • 1
    Electronic Resource
    Electronic Resource
    Springer
    Medical & biological engineering & computing 31 (1993), S. 165-170 
    ISSN: 1741-0444
    Keywords: Breast cancet ; Image quality index ; Monte Carlo ; Time of flight ; Transillumination
    Source: Springer Online Journal Archives 1860-2000
    Topics: Biology , Chemistry and Pharmacology , Medicine
    Notes: Abstract A Monte Carlo simulation has been developed to predict the quality of time-resolved images of the breast by transillumination. The smallest diameter of a detectable carcinoma located in the breast has been computed. The simulation suggests that time-resolved imaging of the breast is possible and invaluable in the near infra-red (NIR) by transillumination. The enhancement of the transfer function by the introduction of time-resolved detection is limited by the contribution of noise at short integration times. The estimated diameter of the smallest detectable sphere is derived from the image quality index (IQI) theory and its value is around 4 mm. The simulated images of an absorbing sphere (approximating the carcinoma) within a homogeneous medium (approximating the surrounding tissue) show a significant improvement of the image with short integration time.