    ISSN: 1612-1112
    Keywords: Gas chromatography ; Modified thermal energy analyzer ; Selective detection of nitrogen compounds
    Source: Springer Online Journal Archives 1860-2000
    Topics: Chemistry and Pharmacology
    Notes: Summary A modification to the TEATM Analyzer is described, which allows it to be used as a highly selective GC detector for nitrogen compounds as well as for N-nitrosamines and nitro compounds. The modified Analyzer is as sensitive as the AFID, but has a much higher selectivity towards nitrogen compounds than the AFID. Its response appears to be molar, with an output that is dependent only on the number of nitrogen atoms present. The performance of the modified Analyzer for the analysis of ammonia, amines,N-nitrosamines, nitriles and organonitro compounds is reported and discussed, as is its application to the detection of these compounds in bacon and beer samples.