Library

Your email was sent successfully. Check your inbox.

An error occurred while sending the email. Please try again.

Proceed reservation?

Export
  • 1
    ISSN: 1612-1112
    Keywords: Supercritical fluid chromatography ; Ion mobility detection ; Mixed mobile phase ; Molecular connectivity
    Source: Springer Online Journal Archives 1860-2000
    Topics: Chemistry and Pharmacology
    Notes: Summary An ion mobility detector (IMD) was evaluated for open tubular column supercritical fluid chromatography (SFC) when organic solvent-modified supercritical CO2 was used as mobile phase. It was found that the SFC/IMD interface design in which the SFC capillaray restrictor was directly inserted into the ionization region of the IMD was not acceptable because of low sensitivity that resulted from the effect of the modifier on detector temperature and mechanism of detection. A new interface utilizing a heated nebulizer gas to provide heat to the restrictor and to minimize the formation of ion clusters, and a bent nozzle for enhancing the ionization efficiency of the solute in the IMD ion source are described. Using 5% acetonitrile in CO2, the minimum detectable quantity (S/N=3) for pyrene was improved from 25.2 ng to 2.1 ng with the new detector design. This compares to a minimum detectable quantity of 0.1 ng when using neat CO2 as mobile phase. The use of molecular connectivity calculations to predict the drift times of selected analytes is also successfully demonstrated.
    Type of Medium: Electronic Resource
    Library Location Call Number Volume/Issue/Year Availability
    BibTip Others were also interested in ...
Close ⊗
This website uses cookies and the analysis tool Matomo. More information can be found here...