    Keywords: Column liquid chromatography ; Limit of detection ; Precision ; Uncertainty prediction
    Source: Springer Online Journal Archives 1860-2000
    Topics: Chemistry and Pharmacology
    Notes: Summary The precision of integration over noisy instrumental output for quantitative analysis is studied. A probability theory is developed to predict the relative standard deviation (RSD) of integration results over an integration domain from one-point integation (peak height measurement) to entire area integration in HPLC. Common integration modes of horizontal zero line and oblique zero line are taken into account, but no peak overlap is assumed. The question of the analytical superiority of peak height measurement or integration for quantitation is answered. In the HPLC apparatus used, the minimum RSD of measurements is found in the integration domain of ca. ±0.5 σ for analytes [peaks are approximated by the Gaussian signal of width, σ (standard deviation)]. The RSD of integration measurements is also shown to depend on the stochastic properties of back-ground noise (uncorrelated noise and correlated 1/f type noise). The theoretical conclusion is verified by Monte Carlo simulation and HPLC experiments for some aromatic compounds.