ISSN:
1612-1112
Keywords:
Gas chromatography
;
UV-Vis detection
;
Gas phase absorption spectrometry
;
Alcohols
Source:
Springer Online Journal Archives 1860-2000
Topics:
Chemistry and Pharmacology
Notes:
Summary This article describes a theoretical/practical study of the parameters which affect the chromatograms obtained when using a molecular diode-array spectrometer as the detector in gas chromatography. The objective is to offer some rules which permit the identification of the optimum signal/noise relation. To achieve this, we study the effect of the different parameters which affect the noise and how to reduce their impact, as well as alternatives for increasing the signal. All the options tested can be applied by correctly programming the spectrometer with BASIC programs, within reach of any use who has even a small understanding of programming. Finally, we consider the effect that the selection of the acquisition conditions may have on the chromatographic resolution. All the studies are carried out using a mixture of alcohols and phenols.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1007/BF02466646