Electronic Resource
Chester
:
International Union of Crystallography (IUCr)
Journal of synchrotron radiation
2 (1995), S. 229-232
ISSN:
1600-5775
Source:
Crystallography Journals Online : IUCR Backfile Archive 1948-2001
Topics:
Geosciences
,
Physics
Notes:
An X-ray mirror fumace is described that has been developed for in situ diffraction experiments at a synchrotron source for temperatures up to 1600 K. It can be used in a reducing as well as in an oxidizing atmosphere. In this kind of furnace, the sample is located at one focus of a rotational ellipsoid and heated by the radiation of a halogen lamp located at the other focus. The furnace has been designed to work on a four-circle diffractometer with minimum restriction of the setting angles.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1107/S0909049595003980