Electronic Resource
[S.l.]
:
American Institute of Physics (AIP)
Review of Scientific Instruments
72 (2001), S. 1677-1679
ISSN:
1089-7623
Source:
AIP Digital Archive
Topics:
Physics
,
Electrical Engineering, Measurement and Control Technology
Notes:
Traditional solid window materials used for x-ray synchrotron beamlines may introduce undesirably high attenuation, or are subject to failure under high heat loads. A plasma window can in principle obviate these problems over a wide range of energies. Experiments were performed at the Brookhaven National Laboratory National Synchrotron Light Source on beamline X6A to study the transmission characteristics of a plasma window using argon as the arc gas. Measurements were made around the Ar K edge and far from resonance. The "white-line" absorption at the K edge was actually suppressed during arc operation as compared to room temperature gas at the same pressure. This is attributed to the high degree of ionization in the plasma. The relative strength of the white line to the edge jump does not seem to be a strong function of arc current at the argon K edge. Away from resonance (∼3 times the edge energy) x-ray attenuation was negligible. © 2001 American Institute of Physics.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1063/1.1344598