Library

Your email was sent successfully. Check your inbox.

An error occurred while sending the email. Please try again.

Proceed reservation?

Export
  • 1
    Electronic Resource
    Electronic Resource
    [S.l.] : American Institute of Physics (AIP)
    Review of Scientific Instruments 60 (1989), S. 1608-1615 
    ISSN: 1089-7623
    Source: AIP Digital Archive
    Topics: Physics , Electrical Engineering, Measurement and Control Technology
    Notes: The development of high-resolution, grazing-incidence soft x-ray monochromators is discussed, and the performance of competing designs is reviewed in terms of efficiency, ultimate resolution, and order sorting. In particular, the characteristics of astigmatically corrected spherical grating monochromator systems (SGM) is examined. A new design is presented in which a variable angle premirror is used to eliminate the wavelength dependent image distance of the SGM and allows the use of order sorting ranges. The properties of a competing design using a plane variable angle premirror, plane grating, and spherical mirror combination are presented, and similarities between this and the corrected SGM are discussed. Flux and resolution data are presented for an undulator monochromator of this latter design.
    Type of Medium: Electronic Resource
    Library Location Call Number Volume/Issue/Year Availability
    BibTip Others were also interested in ...
Close ⊗