ISSN:
1089-7623
Source:
AIP Digital Archive
Topics:
Physics
,
Electrical Engineering, Measurement and Control Technology
Notes:
This paper describes the systematic studies carried out to characterize the controlled clearance piston gauge in the pneumatic pressure region up to 5.09 MPa by measuring its fall rate as a function of the jacket pressure (Pj) using different working fluids such as argon, nitrogen, helium, and hydrogen. It is also observed that a small ambient temperature fluctuation does not affect the overall uncertainty in the measured pressure. A relatively better reproducibility in the stall jacket pressure (Pz) values is obtained for helium and hydrogen as compared to that of argon and nitrogen; therefore it is shown that helium can also be used in addition to the commonly used nitrogen in the pneumatic pressure region. But the use of helium requires a higher value of Pj compared to that of nitrogen in order to get a good floating condition of the gauge. In addition, a detailed study is reported of the effective area of the simple piston gauge with different working fluids used during the calibration.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1063/1.1140026