Library

Your email was sent successfully. Check your inbox.

An error occurred while sending the email. Please try again.

Proceed reservation?

Export
  • 1
    Electronic Resource
    Electronic Resource
    Weinheim : Wiley-Blackwell
    Journal of High Resolution Chromatography 22 (1999), S. 153-158 
    ISSN: 0935-6304
    Keywords: Vapor analysis ; nerve agents ; sulfur mustard ; data acquisition ; automation ; valve injection ; Chemistry ; Analytical Chemistry and Spectroscopy
    Source: Wiley InterScience Backfile Collection 1832-2000
    Topics: Chemistry and Pharmacology
    Notes: ---A method is presented for the continuous analysis of generated vapors of the nerve agents soman and sarin and the blistering agent sulfur mustard. By using a gas sampling valve and a very short (15 cm) column connected to an on-column injector with a “standard length” column, the system can either be calibrated or used for performing high speed gas analyses. When using a flame ionization detector, the detection limit was ca. 100 ppb (ca. 0.5-1.0 mg/m3). This technique is applied in inhalation toxicokinetic studies of nerve agents and mustard gas in the guinea pig.
    Additional Material: 6 Ill.