Library

Your email was sent successfully. Check your inbox.

An error occurred while sending the email. Please try again.

Proceed reservation?

Export
  • 1
    ISSN: 0935-6304
    Keywords: Large volume on-column injection ; partially concurrent solvent evaporation ; solvent trapping ; retaining precolumn ; Chemistry ; Analytical Chemistry and Spectroscopy
    Source: Wiley InterScience Backfile Collection 1832-2000
    Topics: Chemistry and Pharmacology
    Notes: ---When 0.53 mm i.d. uncoated precolumns connected to a solvent vapor exit are used for sample introduction with partially concurrent solvent evaporation, substantial losses of volatile solutes are often observed. They were found to be the consequence of solute accumulation at the front end of the flooded zone, which in turn is the result of a strong pressure drop over the flooded zone owing to the formation of plugs of sample liquid. The pressure drop causes significant solvent evaporation at the front, which enriches the solute material there and causes its loss. The use of 0.32 mm i.d. restrictions between the uncoated precolumn and the vapor exit greatly reduced this problem.
    Additional Material: 10 Ill.
    Type of Medium: Electronic Resource
    Library Location Call Number Volume/Issue/Year Availability
    BibTip Others were also interested in ...
Close ⊗
This website uses cookies and the analysis tool Matomo. More information can be found here...