    Weinheim : Wiley-Blackwell
    Journal of High Resolution Chromatography 22 (1999), S. 509-513 
    ISSN: 0935-6304
    Keywords: Fast GC ; fast temperature programming ; resistive heating techniques ; short columns ; Chemistry ; Analytical Chemistry and Spectroscopy
    Source: Wiley InterScience Backfile Collection 1832-2000
    Topics: Chemistry and Pharmacology
    Notes: ---One possible way to speed up a gas chromatographic analysis is the application of fast temperature programming by using resistive heating techniques. With this heating technique programming rates up to 20° per second can be reached. A relative standard deviation of retention times better than 0.2% is obtained. Using fast temperature programming the analysis-times of a mineral oil sample, an industrial oligomer sample, and toxic compounds in diesel fuel have been reduced 5 to 20 times, compared to a standard temperature programmed analysis. In most cases resistive heating cannot be applied to reduce the analysis time of a complex sample. The use of fast temperature programming is preferable to the use of short columns and columns operated at above-optimum carrier gas velocities.
    Additional Material: 6 Ill.