    Intensive care medicine 21 (1995), S. 753-758 
    ISSN: 1432-1238
    Keywords: Artificial ventilation ; Pediatric intensive care ; Positive end-expiratory pressure
    Source: Springer Online Journal Archives 1860-2000
    Topics: Medicine
    Notes: Abstract Objective To analyze efficiency and reliability of 4 modern neonatal ventilators under difficult test conditions. The ventilators tested were: Babylog 8000 (Dräger Medical), BP 2001 (Bear Medical Systems), Sechrist IV 100 B (Sechrist Industries), Infant Star (Infrasonics INC). Measurements and results Gas flow generation was tested by comparison of preset flow values with no resistance in the circuit to flow values obtained during interposition of a resistance in the inspiratory circuit. A decrease in gas flow was observed when interposition of a resistance in the inspiratory circuit increased peak inspiratory pressure to 60 cmH2O (gas flow decreased by 8% to 24% depending on the ventilator tested). The pressure limiting valve and the positive end-expiratory pressure valve were also evaluated in order to test their behaviour under different flow conditions. Flow-dependence of the pressure was noted for all ventilators except Babylog 8000. Assessment of the reliability of pressure monitoring revealed either ‘under’ or ‘over’ estimation of peak inspiratory pressure and positive end-expiratory pressure depending on the ventilator tested. Conclusion For the best clinical use of mechanical ventilators, neonatologists should be aware of these limitations. Therefore a regular assessment of ventilator performance and monitoring reliability is recommended.
    Type of Medium: Electronic Resource
    Library Location Call Number Volume/Issue/Year Availability
    BibTip Others were also interested in ...
Close ⊗
This website uses cookies and the analysis tool Matomo. More information can be found here...