ISSN:
1089-7623
Source:
AIP Digital Archive
Topics:
Physics
,
Electrical Engineering, Measurement and Control Technology
Notes:
Several improvements in the high-voltage heavy-ion-beam probe (HIBP) are discussed. (1) It is clearly found that the precision slide mount of the detector plates 30° parallel to the base electrode is very effective for the determination of the in-plane entrance angle of the beam in the analyzer to estimate the error in the potential measurement. (2) A two-staged optical trap in the HIBP greatly reduced the effect of the UV radiation in the analyzer. (3) A multiple-plate detector up to 13 measurement points clearly showed the direction of the propagation of the turbulence and path-integral effects. © 1995 American Institute of Physics.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1063/1.1146399