Library

Your email was sent successfully. Check your inbox.

An error occurred while sending the email. Please try again.

Proceed reservation?

Export
  • 1
    Electronic Resource
    Electronic Resource
    [S.l.] : American Institute of Physics (AIP)
    Review of Scientific Instruments 61 (1990), S. 3536-3539 
    ISSN: 1089-7623
    Source: AIP Digital Archive
    Topics: Physics , Electrical Engineering, Measurement and Control Technology
    Notes: This paper describes the in situ calibration of the BF3 detector system used for measuring the total neutron yield from the FTU tokamak. Two different neutron sources, 252Cf (7.23×107 n/s) and AmBe (2.18×106 n/s), were used. The calibration was performed by placing the two sources at different radial and toroidal positions inside the vacuum vessel. Data on the response of the different detectors as a function of the source position are presented. The effect of the different spectra of the neutron sources is pointed out. Finally, the procedure for the computation of the relationship between the local flux at the detectors and the total neutron yield from an extended plasma is presented.
    Type of Medium: Electronic Resource
    Library Location Call Number Volume/Issue/Year Availability
    BibTip Others were also interested in ...
Close ⊗
This website uses cookies and the analysis tool Matomo. More information can be found here...