    Electronic Resource
    Electronic Resource
    Springer
    Medical & biological engineering & computing 38 (2000), S. 49-55 
    ISSN: 1741-0444
    Keywords: Gastric pacemaker ; Serosal activity analysis ; Gastric slow wave
    Source: Springer Online Journal Archives 1860-2000
    Topics: Biology , Chemistry and Pharmacology , Medicine
    Notes: Abstract An algorithm for determining the frequency and propagation time of the gastric slow wave has been designed for integration into a demand gastric pacing system. The algorithm analyses the serosal activity in both the time and frequency domains, and the results are compared to produce a conclusion only when the values are within 5% of each other. Thus, the probability of inappropriate intervention is reduced, at the expense of unidentified segments. The system is verified by comparing the conclusions produced by the algorithm with conclusions from hand analysis of seven canine and one human serosal recordings. The algorithm correctly identifies the slow-wave frequency in the distal portion of the stomach for 90% of the segments, while producing no incorrect results. Slow-wave propagation times in the antrum are correctly identified for 84% of the segments, with no incorrect identifications.
    Type of Medium: Electronic Resource
    Library Location Call Number Volume/Issue/Year Availability
    BibTip Others were also interested in ...