A cold on-column injection system for capillary gas chromatography (GC) applications was constructed. It was based upon a conventional split/splitless capillary GC inlet, which in turn was a modification of a conventional packed GC column inlet. The heart of the laboratory constructed cold on-column inlet design was a disposable pyrex micro-sampling pipet, which functioned as a needle guide for sample injection. The sample was injected through a traditional GC septum. Construction of the injection system is described and applications are illustrated by separations of a variety of complex mixtures.