ISSN:
0935-6304
Keywords:
Trace analysis
;
Capillary chromatography
;
Solvent effect inlet
;
Chemistry
;
Analytical Chemistry and Spectroscopy
Source:
Wiley InterScience Backfile Collection 1832-2000
Topics:
Chemistry and Pharmacology
Notes:
A sample inlet, based on the static solvent effect has been designed and constructed. The inlet can handle samples up to 20 μl which allows for the detection, using an F. I. D., of constituents in the parts per billion range. The reproducibility of the inlet in terms of peak percentage area is around 1% relative standard deviation over a range of chemical types.
Additional Material:
4 Ill.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1002/jhrc.1240100505