Library

Language
Preferred search index
Number of Hits per Page
Default Sort Criterion
Default Sort Ordering
Size of Search History
Default Email Address
Default Export Format
Default Export Encoding
Facet list arrangement
Maximum number of values per filter
Auto Completion
Feed Format
Maximum Number of Items per Feed
feed icon rss

Your email was sent successfully. Check your inbox.

An error occurred while sending the email. Please try again.

Proceed reservation?

Export
  • 1
    ISSN: 1432-1130
    Source: Springer Online Journal Archives 1860-2000
    Topics: Chemistry and Pharmacology
    Notes: Abstract  Chromatographic effluents were on-line analyzed by Zeeman-ETAAS, using a flow-through cell placed in a graphite furnace autosampler as interface. To obtain high sampling rates, the use of fast graphite furnace programmes was studied. Conventional programmes of 96 s were reduced to 18 s by using hot injection (120 °C) and reducing the charring step to 2 s. The increase of the injection volume from 20 to 60 μl lengthened the programme to 46 s. Nickel had to be added to get a comparable response for both inorganic and organic selenium species (selenite and selenomethionine) and to reduce the interferent effect produced in presence of the chromatographic eluent (TRIS 0.01 mol l-1, NH4NO3 0.1 mol l-1, pH 7). The optimized conditions were applied to the speciation of selenium in human erythrocyte lysates by size exclusion LC-ETAAS. Using a high performance size exclusion column selenium could be assigned to proteins of 100 and 35 kDa. Detection limits in the range of 1 ng (2 μg l-1 for 500 μl injection volume) were obtained for the combined technique.
    Type of Medium: Electronic Resource
    Library Location Call Number Volume/Issue/Year Availability
    BibTip Others were also interested in ...
Close ⊗
This website uses cookies and the analysis tool Matomo. More information can be found here...