Electronic Resource
Oxford, UK
:
Blackwell Publishing Ltd
Pacing and clinical electrophysiology
22 (1999), S. 0
ISSN:
1540-8159
Source:
Blackwell Publishing Journal Backfiles 1879-2005
Topics:
Medicine
Notes:
A 72-year-old man with an ICD and a pacemaker was presented with an episode of sustained VT that accelerated to VF. The ICD failed to detect the event and deliver therapy, despite a VT apparently within the VT detection zone. The ICD detected the event after degeneration to VF and delivered appropriate therapy. The high rate event feature of the pacemaker was useful in determining proper function of the ICD along with optimal programming of VT detection.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1111/j.1540-8159.1999.tb00313.x
Permalink
Library |
Location |
Call Number |
Volume/Issue/Year |
Availability |