Abstract We evaluated deuterium oxide (D2O) as a tracer for cardiac output measurements. Cardiac output measurements made by thermodilution were compared with those made by indicator dilution with D2O and indocyanine green as tracers. Five triplicate measurements for each method were made at intervals of 30 minutes in each of 9 anesthetized, mechanically ventilated goats. Cardiac output ranged between 0.68 and 3.79 L/min. The 45 data points yielded a correlation coefficient of 0.948 for the comparison of D2O indicator dilution cardiac output measurements with thermodilution measurements and a linear regression slope of 1.046. D2O indicator dilution measurements were biased by −0.11±0.22 L/min compared with thermodilution measurements and had a standard deviation of ±0.12 L/min for triplicate measurements. Hematocrits ranging between 20 and 50 vol% had no effect on optical density for D2O. D2O is more stable than indocyanine green and approximately one-tenth the price (40 cents per injection compared with $4). The basic instrumentation cost of approximately $9,000 is an additional initial expense, but provides the ability to perform pulmonary extravascular water measurements with a double-indicator dilution technique. D2O has potential as a tracer for the clinical determination of indicator dilution cardiac output measurements and pulmonary extravascular water measurements.