Oxygen consumption was measured by an oxygen sensor after addition of purified fungal amine oxidase to fish extracts. The oxidation of histamine to imidazoleacetaldehyde proceeded stoichiometrically. Based on the equimolar relationship between histamine and oxygen consumption, histamine was determined selectively by the oxygen sensor. Neither sample pretreatment removing interfering materials nor daily calibration by histamine standard was required. Histamine contents in scombroid fish were determined rapidly with good accuracy. AOAC and oxygen sensor methods showed a very high correlation (r=0.999, n=6).