    Keywords: Gas chromatography ; Atomic emission detection ; Aqueous samples ; Solid-phase extraction ; Organophosphorus pesticides
    Source: Springer Online Journal Archives 1860-2000
    Topics: Chemistry and Pharmacology
    Notes: Summary An on-line, solid-phase extraction gas chromatography atomic-emission detection (SPE-GC-AED) system has been set up using an on-column interface to transfer 100 μl of desorbing solvent to the GC part of the system. Analytical characteristics such as recovery, precision and linearity of calibration plots were comparable with those of the off-line combination of SPE-GC-AED using organophosphorus pesticides as test compounds. The fully on-line set-up causes a marked improvement in detection because of the quantitative transfer of the analytes from the SPE module to the GC: detection limits are as low as 5–20 ng l−1 for the analysis of 10 ml raw and spiked surface water samples using the phosphorus channel. Detection levels can be further enhanced by processing up to 100 ml samples. The integrated analytical system is robust. The potential of the on-line set up has been demonstrated for the analysis of surface water and waste water.