ISSN:
1432-1238
Schlagwort(e):
Blood Resistivity
;
Cardiac Output
;
Electrical Impedance
Quelle:
Springer Online Journal Archives 1860-2000
Thema:
Medizin
Notizen:
Abstract A relationship has been established between the haematocrit and the resistivity of whole blood at 37°C for neonates and adults not suffering from renal failure. Values of resistivity obtained in this way were substituted into Kubicek's equation for stroke volume by the electrical impedance technique, the signal pick-up electrodes being placed in standard positions. The calculated cardiac outputs were then compared with those obtained simultaneously from other techniques such as the dye dilution method for adults and the measurement of pulmonary effective capillary blood flow by rebreathing nitrous oxide in neonates. It was found that the impedance method overestimates under these circumstances, and other workers have shown that this is likely to be due to a contribution from the right heart. A haemotocrit-dependent correction factor has been employed in the case of neonates to align the impedance and nitrous oxide results.
Materialart:
Digitale Medien
URL:
http://dx.doi.org/10.1007/BF01683063
Permalink