ISSN:
1741-0444
Schlagwort(e):
Cold-cathode diode
;
Disk cathode
;
Flash X-rays
;
Photon energy control
;
Pulsed X-rays
;
Tube current control
Quelle:
Springer Online Journal Archives 1860-2000
Thema:
Biologie
,
Chemie und Pharmazie
,
Medizin
Notizen:
Abstract Fundamental studies of a repetitive flash X-ray generator using a diskcathode radiation tube are described. The high-voltage pulser employed a modified two-stage surge-Marx circuit. The two condensers in the pulser were charged from 40 to 60 kV, and the electric charges were discharged to the X-ray tube repetitively to generate flash X-rays. The total capacity during the main discharge was 425 pF, and the maximum output voltage from the pulser was about 1·9 times the charged voltage. The flash X-ray tube of the demountable-diode type and was composed of a rodshaped anode tip made of tungsten, a disk cathode made of graphite and a tube body made of polymethylmethacrylate. The peak tube voltage was primarily determined by the anode-cathode (A-C) space, and the peak tube current was less than 0·5 kA. Thus the maximum photon energy could be easily controlled by varying the A-C space, and the tube current roughly increased according to increases in the charged voltage. The pulse width ranged from 40 to 100 ns, and the X-ray intensity was less than 1·1 μC kg−1 at 0·5 m per pulse. The repetition rate was less than 50 Hz, and the effective focal spot size was equivalent to the anode diameter.
Materialart:
Digitale Medien
URL:
http://dx.doi.org/10.1007/BF02446648
Permalink