Schlagwort(e):
Coupled LC-GC
;
Partially concurrent eluent evaporation
;
Evaporation rates
;
Early solvent vapor exit
;
Chemistry
;
Analytical Chemistry and Spectroscopy
Quelle:
Wiley InterScience Backfile Collection 1832-2000
Thema:
Chemie und Pharmazie
Notizen:
Partially concurrent eluent evaporation presupposes an eluent evaporation rate in the GC pre-column that approaches the LC flow rate. Discharging the vapors through the whole GC column, evaporation rates reach 10-30 μl/min, i.e. are suitable just for LC flow rates typical for packed capillary LC columns. With an early vapor exit, evaporation rates are increased to 100-200 μl/min (under extreme conditions to some 800 μl/min), thus fitting the LC flow rates of 2 mm i.d. columns.Evaporation rates were measured for a standard set of pre-columns and conditions. The dependence of the evaporation rate on temperature, inlet pressure, carrier gas, and internal diameter of the retaining pre-column are discussed in order to allow the design of a GC system producing a desired evaporation rate.
Zusätzliches Material:
4 Ill.
Materialart:
Digitale Medien
URL:
http://dx.doi.org/10.1002/jhrc.1240121106