ISSN:
1433-0423
Keywords:
Schlüsselwörter Augeninnendruck
;
Tonometer
;
Kaninchen
;
Keywords Intraocular pressure
;
Tonometer
;
Rabbits
Source:
Springer Online Journal Archives 1860-2000
Topics:
Medicine
Description / Table of Contents:
Abstract Purpose. To assess use of the tonometer ProTon for measuring intraocular pressure in rabbits. Materials and methods. Eleven chinchilla-bastard rabbits were measured under standardized conditions with the applanation tonometer ProTon for 62±15 days. One rabbit was excluded because of intensively defensive reactions. Results. Intraocular pressure in the ten rabbits was 12.15±3.63 mmHg in the right eye and 12.11±3.52 mmHg in the left eye. Conclusion. Intraocular pressure can be measured reliably over a long period with the applanation tonometer ProTon. It is therefore suitable for determining intraocular pressure in experimentally induced glaucoma in rabbits.
Notes:
Zusammenfassung Fragestellung. Untersuchung des lageunabhängigen Applanationstonometers ProTon im Tiermodell. Material und Methoden. Elf Chinchilla-Bastard-Kaninchen wurden unter standardisierten Bedingungen über einen Zeitraum von 62±15 Tagen mit dem ProTon-Applanationstonometer (Fa. Tomey, Deutschland) untersucht. Ergebnisse. An 10 Kaninchen wurden am rechten Auge ein Augeninnendruck von 12,15±3,63 mmHg und am linken Auge ein Augeninnendruck von 12,11±3,52 mmHg gemessen. Ein Kaninchen musste wegen massiver Abwehrreaktionen ausgeschlossen werden. Diskussion. Mit dem Applanationstonometer ProTon konnten über einen längeren Zeitverlauf hinweg zuverlässige Augeninnendruckwerte erhoben werden. Damit eignet es sich zur Druckmessung bei experimentell induzierten Glaukomen bei Kaninchen.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1007/s003470070030
Permalink