ISSN:
1040-7685
Keywords:
microcolumn liquid chromatography
;
reversed-phase mode
;
nonporous C18 columns
;
proteins
;
peptides
;
Chemistry
;
Analytical Chemistry and Spectroscopy
Source:
Wiley InterScience Backfile Collection 1832-2000
Topics:
Chemistry and Pharmacology
Notes:
Proteins and peptides were separated in the reversed-phase mode on microcolumns packed with nonporous octadecyl-group bonded silica gel with an average particle diameter of 4.5 or 20 μm. Separation columns were prepared from glass-lined stainless steel tubing of 39 or 56-mm × 0.5-mm i.d. An artificial mixture of several proteins could be separated within 1 min by split-flow gradient elution. The system was applied to the separation of a triptic digest of mouse amyloid A. On-column sample enrichment allowed direct injection of 2-5 μL of sample solution, leading to improved detectability. The limits of detection were ≤ 1 μg mL-1.
Additional Material:
8 Ill.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1002/mcs.1220030103