ISSN:
0935-6304
Schlagwort(e):
Gas chromatography
;
Trace analysis
;
On-column injection
;
Chemistry
;
Analytical Chemistry and Spectroscopy
Quelle:
Wiley InterScience Backfile Collection 1832-2000
Thema:
Chemie und Pharmazie
Notizen:
Large sample volume on-column injection (up to 250 μL) of n-pentane solutions of halogenated hydrocarbons has been employed for the direct measurement of both low-boiling and high-boiling compounds in what is essentially a single run. Two-bonded phase, fused-silica capillary columns are joined in series, through which the low-boiling compounds are first chromatographed and detected with an electron capture detector. High-boiling compounds are then trapped in a section joining the two columns, and subsequently chromatographed in the second column, using the same detector. This procedure permits analysis at the ppt level.
Zusätzliches Material:
7 Ill.
Materialart:
Digitale Medien
URL:
http://dx.doi.org/10.1002/jhrc.1240070706
Permalink