Electronic Resource
[S.l.]
:
American Institute of Physics (AIP)
Review of Scientific Instruments
57 (1986), S. 225-228
ISSN:
1089-7623
Source:
AIP Digital Archive
Topics:
Physics
,
Electrical Engineering, Measurement and Control Technology
Notes:
The theoretical performance of the hemispherical deflector analyzer has been discussed on the basis of a numerical calculation of electron trajectories by taking the fringing field correctly into account. We optimized both the "Herzog region'' and the electrode sector angle and got foci of electrons which are as good as those for the ideal-field 180 ° analyzer. In the optimized analyzer, even electrons with their energies different from the pass energy are focused well at the exit plate. Moreover, the arriving position of electrons at the exit plate depends linearly upon their incidence energy. The possibility of using a position-sensitive detector is discussed.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1063/1.1138974