ISSN:
0935-6304
Schlagwort(e):
Gas chromatography
;
Capillary, glass
;
Mobile phase flow programming
;
Comparison with temperature programming
;
Short capillaries offer considerable shorter analysis time when flow programmed
;
Chemistry
;
Analytical Chemistry and Spectroscopy
Quelle:
Wiley InterScience Backfile Collection 1832-2000
Thema:
Chemie und Pharmazie
Notizen:
The technique of programming the carrier gas flow rate in gas chromatography, especially in connection with the use of capillary columns shorter than 10 m can significantly accelerate GC analyses. Equations for calculation of the parameters of the exponential flow function and retention data are described. The effects of flow programming in a short capillary column are shown in a few chromatograms. Different programming rates are tested and compared with temperature programming.
Zusätzliches Material:
7 Ill.
Materialart:
Digitale Medien