Library

Your email was sent successfully. Check your inbox.

An error occurred while sending the email. Please try again.

Proceed reservation?

Export
  • 1
    Electronic Resource
    Electronic Resource
    Chichester : Wiley-Blackwell
    Biological Mass Spectrometry 6 (1979), S. 187-190 
    ISSN: 0306-042X
    Keywords: Chemistry ; Analytical Chemistry and Spectroscopy
    Source: Wiley InterScience Backfile Collection 1832-2000
    Topics: Chemistry and Pharmacology
    Notes: Quantification of low levels of psychotropic drugs (10-7 to 10-9 g ml-1) in small volumes of plasma requires sensitive and accurate methods. Validation of these methods is best achieved by comparing results obtained using several techniques. In this study, amphetamine levels in plasma were measured using gas chromatography mass spectrometry and radioimmunoassay. Correlation of the results obtained by the two methods was found to be positive and high (R = 0.9822). The average coefficient of variation between assays for gas chromatography mass spectrometry was 5.8% and for radioimmunoassay was 12.3%, while the average coefficient of variation within assays for gas chromatography mass spectrometry was 4.9% and for radioimmunoassay 6.9%. Although gas chromatography mass spectrometry was 1.9 times more sensitive than radioimmunoassay, for most purposes, the convenience of the radioimmunoassay method outweighs the technical superiority of gas chromatography mass spectrometry.
    Additional Material: 2 Ill.
    Type of Medium: Electronic Resource