Library

Your email was sent successfully. Check your inbox.

An error occurred while sending the email. Please try again.

Proceed reservation?

Export
  • 1
    ISSN: 0306-042X
    Keywords: Chemistry ; Analytical Chemistry and Spectroscopy
    Source: Wiley InterScience Backfile Collection 1832-2000
    Topics: Chemistry and Pharmacology
    Notes: Conditions were established for analyzing as little as 5 pmol of an underivatized peptide delivered in a glycerol sample matrix as a thin film onto a gold-plated copper sample stage and then bombarded with xenon fast atoms. Calibration of the fast atom bombardment high field mass spectrometer and data system was achieved using cesium iodide/glycerol as a reference. Calibration at several accelerating potentials permitted a mass range from 393 to 5941 u. Several factors were examined that contribute to the quality of the mass spectrum: components within the glycerol such as other peptides, alkali salts, acid and reducing agents; the nature of the fast atom gas; concentration of the peptide delivered to the sample stage; and the effect of the sample stage and sample matrix on sensitivity.
    Additional Material: 13 Ill.
    Type of Medium: Electronic Resource
    Library Location Call Number Volume/Issue/Year Availability
    BibTip Others were also interested in ...
Close ⊗
This website uses cookies and the analysis tool Matomo. More information can be found here...