ISSN:
1618-2650
Source:
Springer Online Journal Archives 1860-2000
Topics:
Chemistry and Pharmacology
Notes:
Summary Graphite has been employed as a working electrode in the controlled potential coulometric determination of uranium and plutonium. The couples U(VI)/U(IV) and Pu(IV)/Pu(III) employed for analysis have diverse redox potentials and commonly the working electrodes employed are mercury and platinum. A graphite electrode in the shape of a beaker showed satisfactory performance for the quantitative reduction of U(VI) to U(IV) and Pu(IV) to Pu(III) and also for quantitative oxidation of Pu(III) to Pu(IV). Studies on the levels of the background current, blank values and their reproducible behaviour in acid media have been carried out with a view to achieve good precision and accuracy. A software-based predictive evaluation technique of end-point charge has been investigated. The results have shown that the graphite electrode can be used for the determination of both uranium and plutonium in the presence of each other with a precision and accuracy of better than ±0.5%.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1007/BF00324240