    Fresenius' journal of analytical chemistry 358 (1997), S. 677-682 
    ISSN: 1432-1130
    Source: Springer Online Journal Archives 1860-2000
    Topics: Chemistry and Pharmacology
    Notes: Abstract Well-defined oxygen standard solutions were obtained by the electrolysis of water in a coulometric oxygen generator. The generator was integrated into a flow system that includes the degassing of the carrier electrolyte, the generation of dissolved oxygen and the temperature control of the carrier electrolyte. The current efficiency of oxygen generation was found to be 100% by the Winkler titration method. Calibrations of a home made laboratory sensor and a WTW CellOx dissolved oxygen sensor have been made in a concentration range of 0.02 to 8 mg/L at temperatures from 5°C to 30°C. The calibration of the WTW sensor on water vapour saturated air was compared with the electrochemical calibration method. Both methods gave reliable results provided that the temperature equilibration between the sensor and the ambient air was successful.