Library

Your email was sent successfully. Check your inbox.

An error occurred while sending the email. Please try again.

Proceed reservation?

Export
  • 1
    Electronic Resource
    Electronic Resource
    Copenhagen : International Union of Crystallography (IUCr)
    Applied crystallography online 10 (1977), S. 18-20 
    ISSN: 1600-5767
    Source: Crystallography Journals Online : IUCR Backfile Archive 1948-2001
    Topics: Geosciences , Physics
    Notes: The cooling of single-crystal samples to helium temperatures within a four-circle neutron diffractometer has been achieved by a novel continuous-flow system. The system relies solely on helium for its cooling power with two flow circuits being used. The first delivers liquid to a reservoir cryostat of capacity 1–4 l located close to the Eulerian cradle of the diffractometer and the second draws the refrigerant from this cryostat via a flexible transfer tube into the sample region of a compact cooling unit mounted off the cradle. This arrangement allows unhindered motion of the diffractometer circles for sample temperature operation in the range 3.9–300 K. The system requires a total helium flow of 1.0 l h−1 to maintain a temperature of 20±0.1 K, decreasing to 0.5 l h−1 at higher temperatures.
    Type of Medium: Electronic Resource
    Library Location Call Number Volume/Issue/Year Availability
    BibTip Others were also interested in ...