Construction and performance of a spectrometer based on a high-voltage generator and W-anode X-ray tube are described. W Kα1 radiation is monochromated using a bent Ge (400) crystal. The obtained monochromatic flux is 107 photons s−1 and the beam size at the focal point 15 × 0.5 mm. Comparison is made with a 241Am γ-ray source, which has an almost equal photon energy. The W Kα1 beam does not exhibit the low-energy tail typical for the γ-ray source but, on the other hand, a small contribution of doublet energy was observed. Directional Compton profiles of pyrolytic graphite were measured with both sources and essentially the same features were found.