The design, fabrication, and performance of an apparatus for measurements of rf penetration depth in superconducting samples are described. The instrument makes use of a close-cycle helium refrigerator and is capable of measuring vortex dynamics in superconducting sample of volume as small as 1×10−3 cm3 with a signal-to-noise ratio of 1400. The system has been tested for a tunnel diode oscillator and an IC 74LS04 oscillator in the 3–5 MHz frequency range. Both magnetic field magnitude and angular variations to the resolution of 1 G and 0.05°, respectively, are possible with the apparatus. The system performance is evaluated by measuring the dynamical behavior of vortices in platelet samples of Bi2Sr2CaCu2O8 and (Bi-Pb)2Sr2Ca2Cu3O10 superconductors. © 1999 American Institute of Physics.