    Review of Scientific Instruments 68 (1997), S. 3761-3765 
    ISSN: 1089-7623
    Source: AIP Digital Archive
    Topics: Physics , Electrical Engineering, Measurement and Control Technology
    Notes: A superconducting quantum interference device magnetometer system designed for low field static and dynamic susceptibility measurements is described. Magnetic shields reduce the residual field in the sample space to about 0.04 A/m (0.5 mOe). A highly stable dc magnetic field is generated by a small superconducting solenoid working in the persistent mode. The temperature control system yields a resolution better than 10 μK in the range of 30–100 K. Combining ac susceptibility and dc magnetic relaxation measurements, the system allows dynamic susceptibility studies in a time window ranging from 3×10−6to 106 s. The sensitivity of the system is 10−12 Am2 (10−9 emu). © 1997 American Institute of Physics.