Library

Your email was sent successfully. Check your inbox.

An error occurred while sending the email. Please try again.

Proceed reservation?

Export
  • 1
    Electronic Resource
    Electronic Resource
    [S.l.] : American Institute of Physics (AIP)
    Review of Scientific Instruments 59 (1988), S. 1388-1393 
    ISSN: 1089-7623
    Source: AIP Digital Archive
    Topics: Physics , Electrical Engineering, Measurement and Control Technology
    Notes: A very-high-sensitivity magnetometer is described in detail. The sensitivity of the apparatus is typically of the order of 10−9 A cm2/(Hz)1/2, so that changes in magnetic moment of 10−10 A cm2 can be detected. The magnetic moment can be calculated by simulation of the measurement system. An example of a hysteresis loop is shown. The size of the sample is determined by electron microscopy. The physical limits of the measurement system are worked out and discussed.
    Type of Medium: Electronic Resource
    Library Location Call Number Volume/Issue/Year Availability
    BibTip Others were also interested in ...
Close ⊗
This website uses cookies and the analysis tool Matomo. More information can be found here...