thermal lens
;
vanadium analysis
;
trace analysis
;
spectrophotometry
;
optimization
Source:
Springer Online Journal Archives 1860-2000
Topics:
Chemistry and Pharmacology
Notes:
Abstract Trace level vanadium determination is reported using a dual beam thermal lens spectrometer. The thermal lens was generated using an argon ion beam laser (pump beam) which was focused into a sample cuvette. The thermal lens signal (TLs) was monitored with a He-Ne laser beam and a photodiode detector. Multichannel averager software was developed for processing the transient TLs. The optimal set up, ensuring maximum sensitivity and linear calibration graphs was obtained using experimental design techniques. Under optimized conditions, the detection limits for aqueous and ethanol-water (2+3 v/v) and (4+1 v/v) vanadium complex solutions were, respectively, 0.0071 mg/l, 0.0065 and 0.0039 mg/l.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1007/BF01243045
Permalink