Summary An investigation through interlaboratory comparison using different analytical techniques has been carried out in order to assess the suitability of a plasma protein solution as a source for a trace element reference material in clinical analysis. Reasonable agreement was obtained for a number of elements from the range studied: Al, Ba, Cr, Cu, Fe, Mg, Mn, Mo, Ni, Se, Sr, V and Zn. The techniques used included flame and furnace atomic absorption spectrometry (FAAS and ETA-AAS), furnace atomic emission spectrometry (ETA-AES), inductively coupled plasma atomic emission spectrometry (ICP-AES), inductively coupled plasma mass spectrometry (ICP-MS), and simultaneous multi-element atomic absorption spectrometry (SIMAAC). Results indicated that this plasma protein solution may prove useful as a source for a reference material covering trace element levels outwith the range found in normal human plasma.