Abstract Precise dosimetric calibration of the radiation field of the Gammacell irradiator has been carried out. For measuring spatial dose distribution a cavity ionization chamber in conjuction with a Farmer dosimeter and CaF2 thermoluminescent dosimeters were used. The results make possible the calculation of doses received by samples of different sizes and shapes.